Lenterra Optical Interrogators (LOI)
Lenterra Optical Sensor Interrogator (LOI) is a compact, robust, fast measurement unit designed for the high resolution interrogation of optical sensors (Fiber Bragg Grating or Whispering Gallery Mode sensors) in laboratory or industrial environments.
LOI utilizes fast, low noise Distributed Feedback diode laser with narrow linewidth for full spectrum scanning. Adjustable photodiode gains allow for measurement of returned light intensity in a wide dynamic range. LOI is controlled by any standard PC through USB interface. The unit can be used with third party FBG or WGM sensors as well as with Lenterra’s RealShear™ and DFF sensors for measurements of such parameters as wall shear stress, drag force, temperature, strain, viscosity and others.
LOI-2F
LOI-2F interrogates one sensor, a DFF or RealShearTM, that can be placed in various places inside the processing equipment, for example next to an agitator blade of a mixer. Depending on the flow speed and nature of the material, the sensor with required sensitivity should be selected. A standard or custom adapter can be used, See available adapters here.
LOI-4F
Specifications
| Model | LOI-2F | LOI-4F |
|---|---|---|
| Mechanical, Environmental, Electrical Properties | ||
| Length | 40 cm (15.5") | 47 cm (18.5") |
| Width | 30 cm (11.7″) | |
| Height | 10.5 cm (4.2″) | |
| Weight | 4.7 kg (10.4 lbs) | 6 kg (13.2 lbs) |
| Voltage requirements | 100-240 VAC @ 47-63 Hz | |
| Operating temperature | From 0°C to 40 °C | |
| Storage temperature | From -10 °C to 50 °C | |
| Optical Properties | ||
| Light source | DFB diode laser | |
| Central wavelength | 1540-1560 nm | |
| Linewidth, FWHM | <10 MHz | |
| Scan range: | ||
| via laser temperature1 | 3.7 nm | |
| via laser current1, 2 | 1.2 nm | |
| Scan rate: | ||
| via laser temperature1 | 0.1 nm/s | |
| via laser current1 | from 0.33 nm/s to 650 nm/s | |
| Scan frequency (via laser current)3 | from 0.25 Hz to 500 Hz | |
| Wavelength increment1 | ||
| at 0.25 Hz | 0.8 pm | |
| at 10 Hz | 0.8 pm | |
| at 100 Hz | 0.7 pm | 1.6 pm |
| at 500 Hz | 2.9 pm | 5.4 pm |
| FBG spectrum tracking repeatability1,4 | ||
| at 0.25 Hz | 0.5 pm | |
| at 10 Hz | 0.3 pm | |
| at 100 Hz | 0.2 pm | |
| at 500 Hz | 0.4 pm | |
| Number of optical channels | 2 | 4 |
| Connector | Duplex LC/PC | Two FC/APC |
| Photodiode amplifier gain | 105, 106, 107 V/A | |
| Data Processing Capabilities | ||
| Interface | USB (with host computer) | |
| Software type | LabVIEW based data acquisition and control software | |
| Software capabilities | Spectra recording, spectra tracking, data logging and processing, instrument control | |
| Integration | Optional analog output Optional sync output TTL pulse |
|